|
Untitled Document
CTS
Home > Laboratory
Exercises
> Microscope
Use
Use
of the Light Microscope
It is essential that you
familiarize yourself with the operation of the light
microscope provided to you in order to maximize your success in the laboratory.
Please systematically work your way through the material presented here.
Additional information on the light microscope is provided in the course
textbook [Gardner and Hiatt, p. 3]. If you have any questions about your
understanding of this material or the use of any of the components of your
microscope, please seek assistance from one of the lab instructors. Your
understanding and proper use of the light microscope is critical to success
in the laboratory component of this course.
Locate and handle the various parts of your microscope and be sure
that you know their functions. In particular, make sure that you
can locate and use the field diaphragm, the condenser diaphragm, the
condenser centering knobs,
stage controls, and the condenser focusing knob.
Each time you use the microscope, you should set it up for proper ("Koehler")
illumination. You will also need to adjust the eyepieces for comfortable viewing
with both eyes.
Use the following steps to adjust the eyepieces, establish Koehler illumination,
and acquaint yourself with use of the several objectives:
- Always move the microscope
with two hands; you will have to tip the microscope to remove it from the
cabinet. You may have to attach the black power cord (found behind the
scope in the cabinet). Set the microscope on the counter top with its single
foot and the binocular viewing body towards you. Turn on the lamp with
the amber switch in the right rear foot and adjust its intensity with the
black circular knob in the base (on the left). Place any slide from your
set on the microscope stage. Swing the 4x (scanning) objective into the
optical axis, focus the image of the slide using the focus knobs (labeled "1" in diagram),
and then switch to the 10x objective and refocus.
- Match the eyepieces
of the microscope to the distance between your eyes by rotating the two
parts of the binocular body inward or outward. Compensate for any difference
in visual acuity between your two eyes by first blocking vision through
the eyepiece with the knurled ring and focusing the microscope objective
through the other eyepiece. Then, without changing the focus of the microscope
objective, look at the slide through the knurl-ringed eyepiece and obtain
the sharpest image by rotating the ring on that eyepiece. You may need
to repeat these adjustments each time you use the microscope.
- Now, close the FIELD diaphragm
of the microscope (labeled as #4 in the diagram,
large knurled ring around the light exit window in the microscope base)
and focus the condenser by turning the black knurled knob on the
left side of the microscope until the smallest circle of light is observed
and the edge of the field diaphragm is in sharpest focus. This will also
produce the least colorful range of the fringes around the diaphragm
image.
- Open your field
diaphragm until it approaches the edge of the visual field and center
it by using the two small knurled screws on the front sides of the
condenser (# 6 in the diagram). Then
open the field diaphragm until the entire field of view is evenly
illuminated.
- Next,
fully open the APERTURE diaphragm on the condenser
(large knurled ring in plane of item 7 in the diagram)
so that the maximum amount of light comes through. Now
close this diaphragm slowly until you can just detect an
effect upon the visual field. This should provide the proper
aperture adjustment for this lens. Note that you should NOT
attempt to control the brightness of the illumination with
the condenser aperture diaphragm; you may use it, however,
to change contrast in the image. Closing the condenser aperture
diaphragm will eliminate stray light and increase the contrast
in the image. Overall brightness should be controlled by
adjusting the light rheostat, using the black circular knob
within the left side of the base.
These
several steps to focus the condenser and adjust the field
and aperture diaphragms will produce light best adjusted
for proper microscope use, identified as "Koehler illumination".
- Change
to your high-dry (40x) lens and readjust
both field and aperture diaphragms as necessary
to reproduce Koehler illumination with this
objective. As you will observe, when the
objective is changed, reestablishment of
proper illumination requires slight changes
in field and aperture diaphragms as well.
- Now
practice
use of the
oil immersion
(100x) lens.
Turn the
objective
nosepiece
half-way
between the
high-dry
and oil immersion
lenses and STOP.
Place a small
drop of
immersion
oil on the
slide over
the spot
of light
from the
condenser.
Swing the
oil immersion
lens into
the oil droplet,
note that
the lens "picks
up" the drop
of oil, and
focus the
lens on the
slide. This
lens cannot
be focused
sharply unless
its tip is
immersed
in oil. Be
careful to
avoid getting
oil by mistake
on any of
the other
objectives,
especially
the high-dry
(40x) lens,
and do not
clean any
of the lenses
with the
solvent in
the brown
glass bottles.
- When
you have
finished viewing
with the
100x lens,
lower the
stage to
break the
connection with
the oil
droplet and
gently but
thoroughly wipe
all oil
from the
lens and
the stage
using lens
paper. The
same cleaning
procedure can
be used
if you
accidentally get
immersion oil
on any
other objective
lens. Lenses
can be
irreparably damaged
if significant
amounts of
oil are
inadvertently left
on them
after use. Clean
the slide
with lens
paper and
the solvent
provided in
the brown
glass bottles.
Use the
100X oil
immersion lens
sparingly.
Summary:
Whenever you first use the microscope or when you change objectives, do a
realignment of the microscope: 1) focus on the specimen, 2) close the field
diaphragm, 3) adjust position of condenser lens to obtain a sharp image of
the field diaphragm, 4) open the field diaphragm, 5) adjust contrast using
the condenser aperture diaphragm.
|